Search

EM Test CTR2

Calibration Target
  • Current target as per IEC/EN 61000-4-2
  • 2 Ohm impedance +/- 5%
  • Test voltage up to 30kV
  • Insertion loss +/- 0.5dB up to 1GHz
  • Insertion loss +/- 1.2dB up to 4GHz
  • Includes 20dB attenuator and 1m cable RG400
  • 50ohm Adapter line CTR 2-AD available (option)

Quick Enquiry

If you want to know more about EM Test CTR2
Calibration Target

The CTR 2 is a coaxial current target to monitor Electro Static Discharges built as required by EN/IEC 61000-4-2 Ed.2 standard. The target can be purchased as an option to the ESD generators, model dito or ESD 30. It can be used for measurements up to 30kV. Additional attenuators are available to match the output signal with the input capability of the oscilloscope.

  •  EN 61000-4-2
  •  IEC 61000-4-2
  •  ISO 10605

CTR2/CTR2-AD  Datasheet