Search

Ametek TVT 1-250-16.1

Automatic tapped transformer 40/70/80% 16A
  • Tapped transformer in a compact size
  • Fixed taps at 40 %, 70 % and 80 % of nominal voltage
  • Suitable for 100 V, 115 V and 230 V mains supplies
  • Nominal current 16 A, short-term up to 40 A for 5 seconds
  • Automatic tap selection
  • Compliant to IEC 61000-4-11

Quick Enquiry

If you want to know more about Ametek TVT 1-250-16.1
Automatic tapped transformer 40/70/80% 16A

TVT 1-250-16.1 – tapped transformers for testing voltage dips and interruptions Voltage dips and short interruptions simulate a reduction of the power supply voltage for a certain period of time. Four different test levels are required as per IEC/EN 61000-4-11. The TVT 1-250-xx series of tapped transformer are used to simulate power supply failures such as voltage dips and interruptions providing taps with 40 %, 70 %, 80 % and a direct output with 100 % of the nominal mains supply voltage. Automatic tap selection by means of the 0-10 V DC signal provided by the test generator, max 16 A.

Datasheet TVT 1-250-XX SERIES