EM Test PFS 503Nx.2 Series

3-phase 690Vac

PFS 503Nx.2  Series – Simulator for dips, short interruptions and voltage variations

  • Standalone test generator for dips, short interruptions and voltage variation
  • Full-compliant three-phase system as per IEC 61000-4-11, Ed.2:2004 and IEC 61000-4-34 for testing delta and star supply systems
  • 6 electronic switches, short-circuit protected
  • Rated voltage up to 3x690V AC, 600V DC
  • Nominal current up to 100A per phase
  • DC signal to control external variac
  • Front panel operation
  • USB and GPIB interfaces

Quick Enquiry

If you want to know more about EM Test PFS 503Nx.2 Series
3-phase 690Vac

Unlike any other Power Fail simulator the EM TEST PFS 503N Series represents a real three-phase Power Fail simulator. Not only phase-by-phase testing is possible but any combination of phases and all three phases can be simultaneously tested. Herefore the PFS503N Series is equipped with six individually controlled semiconductor switches. Each switch is electronically short-circuit protected. Additionally the PFS 503N Series is available for nominal currents of up to 100A which is another outstanding and unreached feature for Power Fail testing. The PFS 503N Series offers BNC outputs for individual current monitoring for each phase by means of an oscilloscope. The inrush current can be measured using these BNC outputs. An analog DC signal of 0 to 10V is supplied to drive an external motorised three-phase variac. The PFS 503 can also be used for single phase testing offering a separate DUT output for this purpose.

  • EN 61000-4-11
  • EN 61000-4-29
  • EN 61000-4-34
  • EN 61000-6-1
  • EN 61000-6-2
  • IEC 60601-1-2
  • IEC 61000-4-11
  • IEC 61000-4-29
  • IEC 61000-4-34
  • IEC 61326
  • ECE-R10

PFS 503N Series – The only real three-phase Power Fail simulator

Unlike any other Power Fail simulator the EM TEST PFS 503N Series represents a real three-phase Power Fail simulator. Not only phase-by-phase testing is possible but any combination of phases and all three phases can be simultaneously tested. Herefore the PFS503N Series is equipped with six individually controlled semiconductor switches. Each switch is electronically short-circuit protected. Additionally the PFS 503N Series is available for nominal currents of up to 100A which is another outstanding and unreached feature for Power Fail testing.
The PFS 503N Series offers BNC outputs for individual current monitoring for each phase by means of an oscilloscope. The inrush current can be measured using these BNC outputs. An analog DC signal of 0 to 10V is supplied to drive an external motorised three-phase variac.
The PFS 503 can also be used for single phase testing offering a separate DUT output for this purpose.

Easy to operate

Front panel menu and function keys enable the user to program his test routines quickly and accurately. The cursor allows fast control of all test parameters of the programmed routine, thus test procedures are simplified and confidence is generated that every step is carried out correctly.

iec.control Software for control and documentation

Outstanding user convenience, clearly structured windows and operation features, EM TEST’s comprehensive standards library along with the flexibility to easily generate user specific test sequences are the main features of iec.control software. The software will be automatically configured in accordance with the connected EM TEST generators. Extensive reporting capabilities help the user to create test reports that meet international requirements.
The iec.control is supported by Windows 7, Windows 8 and Windows 10. Remote control is achieved either via Ethernet or optical interface with USB connector at the PC side.
The iec.control supports various interfaces for the communication to external measuring devices.

V3P40xxDS – Tapped autotransformer for testing three-phase dips and interruptions

The tapped autotransformers of the V3P40xxDS series provide the fixed voltage levels as per IEC 61000-4-11 Ed.2:2004 and IEC 61000-4-34. These transformers can easily be configured for both delta connection and star connection to fully cover and comply to the test requirements as per IEC 61000-4-11, Ed.2:2004.

MV3P40xxDS – Motorised variac for three-phase Power Fail tests

A motorised variac is required to set the reduced voltage levels to perform voltage dips as per IEC 61000-4-11 Ed.2:2004 and IEC 61000-4-34. The voltage level is set by means of an analog output signal provided by the Power Fail simulator to any level between 0V and the maximum voltage of the variac. The variac solution also provides capability to perform overvoltage tests.

SVP CL 1700 – Calibration box 1700uF for inrush current verification

SVP CL 100 – Low inductive 100 Ohm carbon calibration load resistor

EM Test PFS 503N Series Datasheet